Chapter 13-8
value between 0% and 100%. The default value is 16%.
—— AGT :Select the concentration of anesthetic agent. It can be set to a
value between 0.0% and 20.0%. The default value is 0.0%.
——
Balance Gas
:Select the type of balance gas, the options are
Air
,
N
2
OandHELIUM. When the most proportions of the mixture is
air, selectAir;When the most proportions of the mixture is N
2
O,select
N
2
O
. When the most proportions of the mixture is Helium, select
HELIUM.
Warning: Please set the CO
2
corrections according to actual
situation, otherwise, the measured value may be
inaccurate and away from actual value.
Setting Wave Scale
Select
Scale
in the CO
2
waveform setting menu, and you can adjust the
position of wave scale manually, and the waveform amplitude will vary along
with it.
Setting Wave Speed
SelectWave Speedin the CO
2
waveform setting menu, and select the
wave speed in the options as required.
13.5 Zeroing
Mainstream CO
2
module and Microstream CO
2
module
Zeroing allows the CO
2
sensor to adjust to the optical characteristics, in
order to obtain accurate readings. While zeroing is recommended the first time
a CO
2
sensor is connected to the monitor, it is only absolutely necessary when
the messageZero Requiredis displayed.
Follow these steps:
1. Ensure that the nasal cannula or airway adapter is not connected to the
patient or close to any source of CO
2
(including the patient's, your own,
exhaled breath and ventilator exhaust valves).
2. SelectZeroin the CO
2
parameter setting menu, this will start zeroing. The
monitor zeroes the sensor and displays the message Zero In Progress for
about 15-20s on the screen. The message disappears upon completion of the
zeroing.