Patient monitor User’s manual
Chapter 9-9
Setting SmartTone
【SmartTone】is a feature that affects pulse beep and can be selected in
the SpO
2
parameter setting menu. When you set it to【On】 , it will allow the
audible pulse beep to beep when the pleth shows signs of motion. The pulse
beep is suppressed during signs of motion when SmartTone is set to
【
Off
】
.
Setting SpHb mode
While monitoring Hemoglobin levels, there are two blood sample sources
from which Hemoglobin readings can be obtained: arterial and venous. You
can select 【 SpHb mode 】 in the SpO
2
parameter setting menu, and you can
select【Arterial】or【Venous】.
Setting SpHb Average
You can select the averaging mode for SpHb value, select【SpHb Average
Time 】 in the SpO2 parameter setting menu, and you can select 【 Short 】 ,
【
Medium
】
or
【
Long
】
.
Setting SpHb Precision
You can select the SpHb precision to be displayed on the screen, select
【 SpHb Precise 】 in the SpO2 parameter setting menu, and select a proper
time as required.
Setting SpHb Unit
You can select【SpHb Unit】in the SpO
2
parameter setting menu, and select
the unit as required. The options are
【
g/dL
】
and
【
mmol/L
】
.
The unit “g”and “ m ” displayed on the screen are short for “g/dL”
“mmol/L”.
Setting Waveform Mode
You can select
【
Wave Mode
】
in the SpO
2
parameter setting menu, and
select【Resp. out】or【Resp. in】.
View the Version of Firmware
You can select【Version】in the SpO
2
parameter setting menu, and you can
view the version of firmware.
9.6 Measurement Limitations
If you doubt the SpO
2
measurements, check the patient’s vital signs first,