Patient monitor User’s manual
Chapter 3-10
with a Temp module (module A), and the module uses the label Tskin, then if
you plug in another Temp module (module B) with the label Tskin, thus the
monitor will close the module B automatically.
To resolve conflict of labels:
—— You can keep on using module A to measure Tskin;
—— If you decide to measure Tskin with module B, you need to modify the
label Tskin with the module A first, and activate the module B once
The monitor can simultaneously support three independent IBP modules
and IBP module in EMS, also support three independent Temp modules and
the Temp module in EMS, but only support one measurement module of other
type, otherwise the monitor will prompt “If you want to activate this
module,please close other activating XX module.”
Take the CO
2
module as an example: a monitor has been equipped with
one CO
2
module (module A), furthermore, the module A is activate, and at that
time, if you plug in another CO
2
module (module B), the module B is defult as
deactivate.If you want to activate module B, the monitor will prompt “If you
want to activate this module, please close other activating CO2 module.”
——You can keep on using module A;
——If you need to use the module B, you must pull out both module A and B,
then plug in the module B. Or you can deactivate the module A and activate
the module B in the【Module Config】window.
The monitor supports USB-interface mouse and allows hot plugging and
While using mouse, you should take care of the following:
The left key of a mouse is the major key, and the right key of the mouse is
To click the major key of a mouse is equivalent to the operation of