Appendix C Alarm messages
C.1 Monitor Alarm Messages
System
Technical alarm:
Alarm messages
Cause
Level
Battery low
Voltage of battery is too low
Medium
Keyboard error
Keyboard error
Low
Record Out of Paper
No paper in the recorder when
recording or the recorder door is open
or recorder is absent.
ECG
Physiological alarm:
Alarm messages
Cause
Level
HR high
HR measuring value is above the high
alarm limit
Medium
User-Selectable
HR low
HR measuring value is below the low
alarm limit
ECG signal too weak
Can’t detect the patient’s ECG signal.
High
ST high
ST measuring value is above the high
alarm limit
Low
User-Selectable
ST low
ST measuring value is below the low
alarm limit
Technical alarm:
Alarm messages
Cause
Level
RALALLV1V2
V3V4V5V6 off
ECG electrode fall off the skin or
ECG cables fall off the monitor
Medium
ECG leads off
Key-leads are disconnected, so it
can’t measure normal ECG
waveform.
V1V2
V3V4V5V6 polarize
ECG electrode polarized
Low
ECG communication error
ECG measurement failure or
communication failure
Medium
ECG noise
ECG noise is too much.
Low
Resp