Patient monitor Users manual
48
Physiological alarm:
Alarm messages
Cause
Level
RR high
RR measuring value is above the high
alarm limit
Medium
User-Selectable
RR low
RR measuring value is below the low
alarm limit
Resp apnea
No signal for breath in specific
interval
High
SpO
2
Physiological alarm:
Alarm messages
Cause
Level
SpO
2
high
SpO
2
measuring value is above the
high alarm limit
Medium
User-Selectable
SpO
2
low
SpO
2
measuring value is below the
low alarm limit
Medium
User can select
Medium or
High
SpO
2
Desat
SpO
2
measuring value is too low.
High
PR high
PR measuring value is above the high
alarm limit
Medium
User-Selectable
PR low
PR measuring value is below the low
alarm limit
SpO
2
pulse timeout
SpO
2
signal is predominantly invalid
and therefore cannot be analyzed
High
Technical alarm:
Alarm messages
Cause
Level
SpO
2
sensor off
SpO
2
sensor may be disconnected
from the patient or the monitor
Low
SpO
2
sensor error
SpO
2
sensor failure
SpO
2
communication error
SpO
2
measurement failure or
communication error
No alarm
Prompt messages:(displayed in SpO
2
parameter area)
Messages
Cause
Level
Search pulse
SpO
2
module is searching for pulse
No alarm
Motion interference
Patient movement too much.
Probe disconnected
SpO
2
probe isnt connected to the
monitor
Technical alarm (Masimo SpO2 module):