Patient monitor User’s manual
Chapter 16-3
3
)
Select the C.O. parameter zone to enter into the menu of
【
C.O.
Setup】, and select the 【C.O. Measure 】 to enter into the measurement
4
)
If the
【
Measure Mode
】
in the
【
Setup C.O.
】
is set as
【
Single
】
,
when the words “Ready for New measurement” are displayed on the
screen, select the 【Start】 to start the C.O. measurement. When there is
prompt information “Inject now…” on the screen, please inject the
solution to the right atrium port of the Swan-Ganz catheter. The optimal
injection rate is 2.5ml/s. The cardiac output, cardiac index and
corresponding curve will display in the measurement window in real time.
5) If the 【 Measure Mode 】 in the 【 Setup C.O. 】 is set as
【 Continuous 】 , select the 【 Start 】 to automatically enter into the
continuous measure mode. When there is prompt information “Inject
now…” on the screen, please inject the solution to the right atrium port of
the Swan-Ganz catheter. The optimal injection rate is 2.5ml/s. The cardiac
output, cardiac index and corresponding curve will display in the
measurement window in real time. When the first measurement ends, the
words “Ready for New measurement, please wait….” are displayed on the
screen, and then a new measurement can be started. When there is prompt
information “Inject now…”on the screen, please repeat the injection
process. When the number of measurements meets the demand, select the
【
Stop
】
to stop the measurement.
Select the C.O. parameter zone to open the menu of 【C.O. Setup】, and
set the items in the menu.
Select the 【C.O. Measure】 to open the C.O. measurement window as
shown in the following figure, please pay attention that whether the TB cable
is well connected and the value of C.O. setting is correspond with actual value