Patient monitor User’s manual
Chapter 16-4
Below the C.O. measurement window, the following function buttons are
included:
1)
Start
: Start a C.O. measurement.
2) Delete : Selectively delete the measured values in the historical
measurement window.
3)
Hemodynamic
: Open the menu of
Hemodynamic
. See 1.5.4 of this
chapter for the detailed content of the menu.
4) C.O. ITEM: Open the C.O. item to browse the measurement results.
5)
Setup Scale
: Open the menu of scale setting to adjust the scale range.
6) Setup C.O.: Open the menu ofSetup C.O.. See 1.5.2 of this chapter
for the detailed content of the menu.
7)
Record
: Record the latest measurement curve and measurement result.
16.5.2 C.O. Setting
Select the Setup C.O. to set the items in the C.O. setting window in
proper order.
1
Measure Mode
:
Single
mode or
Continuous
mode is optional.
2 TI Source : The temperature source of injectate can be input by the
user or measured by the injectate temperature probe. The options include
Manual
and
Auto
. When
Auto
is set, the temperature of injectate
can be obtained through the In-line Injectate sensor cable in real time.
3 Setup TI: When theTI Sourceis automatic, the TI is unadjustable;
when the
TI Source
is manual, the TI is adjustable.
4 Inj.Vol: 3cc, 5cc or 10cc is optional.
5 Cath.Const : The set value of the constant can be obtained in the
Prompt
information
zone
Historical
measurement
window
Currently
measured
C.O. curve
Function
button
C.O.
parameter
zone