Patient monitor User’s manual
Chapter 18-4
An alternation in the waveform scale only alters the appearance of the
waveform rather than influences the signal analysis of the monitor nor the
signal printed in records or reports.
Wave Select
Select Parameter Option.Open the 【 BIS Setup 】 menu. Select 【 Wave
Select】. Three modes can be selected: 【EEG】, 【Trend】 and 【EEG Wave
& Tr.
】
. You can set the certain mode as required.
1) 【EEG】: Display EEG waveform only;
2) 【Trend】: Display trends only;
3) 【 EEG Wave & Tr. 】 : Display EEG waveforms and trends
simultaneously.
18.5.3 BIS parameter Alarm setting
Select Parameter Option. Open the【BIS Setup】 menu,select the【Alarm
Setup】.
Alarm Switch
Select the
【
Alm Limit Setup
】
of
【
Alm Setup
】
menu
,
then you can set
the BIS alarm settings, select the【Alm Switch】 to On or Off , Alarm ON or
OFF can be set.
Alarm Level
Select the【Alm Limit Setup】of 【Alm Setup】menu,then you can set
the BIS alarm settings, select the【Alm Level】, Three options are offered for
needs: Low/Med/High.
Alarm Print
Select Parameter Option. Open the【BIS Setup】 menu,you can select the
【
Alarm Print
】
to On or Off.