Patient monitor User’s manual
Chapter 18-4
An alternation in the waveform scale only alters the appearance of the
waveform rather than influences the signal analysis of the monitor nor the
signal printed in records or reports.
Wave Select
Select Parameter Option.Open the BIS Setup menu. Select Wave
Select. Three modes can be selected: EEG, Trend and EEG Wave
& Tr.
. You can set the certain mode as required.
1) EEG: Display EEG waveform only;
2) Trend: Display trends only;
3) EEG Wave & Tr. : Display EEG waveforms and trends
simultaneously.
18.5.3 BIS parameter Alarm setting
Select Parameter Option. Open theBIS Setup menu,select theAlarm
Setup.
Alarm Switch
Select the
Alm Limit Setup
of
Alm Setup
menu
then you can set
the BIS alarm settings, select theAlm Switch to On or Off , Alarm ON or
OFF can be set.
Alarm Level
Select theAlm Limit Setupof Alm Setupmenuthen you can set
the BIS alarm settings, select theAlm Level, Three options are offered for
needs: Low/Med/High.
Alarm Print
Select Parameter Option. Open theBIS Setup menu,you can select the
Alarm Print
to On or Off.