Patient monitor User’s manual
Chapter 19-24
19.6.2 Set Parameter Alarm
The setting way of turning on /off, level and limit of each parameter alarm
is the same. Therefore, ECG parameter alarm settings are listed here for
example.
1. Turning on\off Alarm
You can set the alarm switch as the following two ways:
Enter the main menu, select
【
Alarm Setup
】
→
【
HR
】
→
【
Alarm Switch
】
,
select【On】or【Off】to turn on or off the alarm.
a. Enter the main menu, select 【Module Setup】→【ECG Setup】→
【
Alarm Switch
】
, select
【
On
】
or
【
Off
】
to turn on or off the alarm.
2. Setting Alarm Level
You can set the alarm switch as the following two ways:
a. Enter the main menu, select
【
Alarm Setup
】
→
【
HR
】
→
【
Alarm Class
】
,
select the level you need.
b. Enter the main menu, select【Module Setup】→【ECG Setup】→【Alarm
Class
】
, select the level you need.
3. Setting Alarm Limit
You can set each parameter’s alarm limit by the following two ways:
a. Enter the main menu, select【Alarm Setup】→【HR】→【Up Limit】
or 【Down Limit】,the alarm setup window appears. Please input the uplimit
or downlimit value. Pay attention that the uplimit shouldn’t be lower than the
downlimit.
b. Through【Module Setup】to Set Alarm Limit
Enter the main menu, select
【
Module Setup
】
→
【
ECG Setup
】
→
【
HR
Up Limit】, you can set the HR high limit. Select【HR Down Limit】, you can
set the HR low limit. Pay attention that the uplimit shouldn’t be lower than the
downlimit.
It is important to set physiological alarm limits properly. The monitor
can’t give medicinal alarm prompt in clinical application with improper
setting of physiological alarm limit.
Warning: Medical personnel should set the alarm limits of
parameters in line with the clinical environment and
existing clinical experience.