Patient monitor User’s manual
Chapter 19-24
19.6.2 Set Parameter Alarm
The setting way of turning on /off, level and limit of each parameter alarm
is the same. Therefore, ECG parameter alarm settings are listed here for
example.
1. Turning on\off Alarm
You can set the alarm switch as the following two ways:
Enter the main menu, select
Alarm Setup
HR
Alarm Switch
,
selectOnorOffto turn on or off the alarm.
a. Enter the main menu, select Module SetupECG Setup
Alarm Switch
, select
On
or
Off
to turn on or off the alarm.
2. Setting Alarm Level
You can set the alarm switch as the following two ways:
a. Enter the main menu, select
Alarm Setup
HR
Alarm Class
,
select the level you need.
b. Enter the main menu, selectModule SetupECG SetupAlarm
Class
, select the level you need.
3. Setting Alarm Limit
You can set each parameters alarm limit by the following two ways:
a. Enter the main menu, selectAlarm SetupHRUp Limit
or Down Limitthe alarm setup window appears. Please input the uplimit
or downlimit value. Pay attention that the uplimit shouldn’t be lower than the
downlimit.
b. ThroughModule Setupto Set Alarm Limit
Enter the main menu, select
Module Setup
ECG Setup
HR
Up Limit, you can set the HR high limit. SelectHR Down Limit, you can
set the HR low limit. Pay attention that the uplimit shouldnt be lower than the
downlimit.
It is important to set physiological alarm limits properly. The monitor
can’t give medicinal alarm prompt in clinical application with improper
setting of physiological alarm limit.
Warning: Medical personnel should set the alarm limits of
parameters in line with the clinical environment and
existing clinical experience.