Patient monitor User’s manual
25
IRMA AX+ sensor:
Gas or vapor
Gas level
Effect (CO
2
)
Effect
(Agents)
Effect
(N
2
O)
N
2
O
60%
HAL
4%
ENF, ISO, SEV
5%
DES
15%
Xe (Xenon)
80%
-10% of reading
1)
He (Helium)
50%
-6% of reading
1)
Metered dose inhaler
Propellants
Not for use with metered dose inhaler propellants
C
2
H
5
OH (Ethanol)
0.3%
C
3
H
7
OH
(Isopropanol)
0.5%
CH
3
COCH
3
(Acetone)
1%
CH
4
(Methane)
3%
CO (Carbon
monoxide)
1%
NO (Nitrogen
monoxide)
0.02%
O
2
100%
Note 1: Interference at indicated gas level. For example, 50 % Helium typically
decreases the CO2 readings by 6%. This means that if measuring on a mixture
containing 5.0 % CO2 and 50 % Helium, the actual measured CO2 concentration will
typically be (1-0.06) * 5.0 % = 4.7 % CO2.
Sidestream AG module (ISA)
Measurement mode
Sidestream
Sampling flow rate
50ml/min ± 10ml/min
Measurement
parameters
CO
2
N
2
OO
2
agent (ISOENFSEVHALDES)
Compensations
Automatic compensation for pressure, temperature
and broadening effects on CO
2
.
Calibration
No span calibration is required for the IR bench. An
automatic zero reference calibration is performed at
startup and then every 24 hours.