Patient monitor Users manual
58
ECG Leads off
Key-leads are disconnected, so it
can’t measure normal ECG
waveform.
Sat (
V1
V2
V3
V4
V5
V6)
ECG electrode polarized
Low
Resp
Physiological alarm:
Alarm message
Cause
Level
RR high
RR measuring value is above the high
alarm limit
Medium
User-selectable
RR low
RR measuring value is below the low
alarm limit
Resp apnea
No signal for breath in specific interval
High
SpO
2
Physiological alarm:
Alarm message
Cause
Level
SpO
2
high
SpO
2
measuring value is above the
high alarm limit
Medium
User-selectable
SpO
2
low
SpO
2
measuring value is below the low
alarm limit
Medium
User can select
Medium or
High
PR high
PR measuring value is above the high
alarm limit
Medium
User-selectable
PR low
PR measuring value is below the low
alarm limit
SpO
2
no pulse
The pulse of patient is too weak and the
system can’t analyze.
High
SpO
2
ultra low
SpO
2
measuring value is too low.
Technical alarm:
Alarm message
Cause
Level
SpO
2
off
SpO
2
sensor may be disconnected
from the patient or the monitor
Low
SpO
2
failure
There is noise or patient movement
too much.
Prompt messages:
Messages
Cause
Level