4 Load Preparation
Instrument mats other than STERRAD Instrument Mats.
Instrument trays other than STERRAD Instrument Trays or
APTIMAX® Instrument Trays.
Implants for which the manufacturer has not specifically recommended
sterilization
in
the STERRAD® NX Sterilizer.
Guidelines for Preparing
Items
to
Be
Sterilized
4 Note:
All
items must be cleaned, rinsed, and thoroughly
dried before being
placed
in
the STERRAD
NX
Sterilizer. Loads containing moisture may cause
cycle cancellations.
Cleaning, Rinsing, and Drying
Cleaning and sterilization are two separate processes. As with all
sterilization methods, proper cleaning
of
instruments and devices
is
a
critical and necessary step prior to sterilization. All items must
be
cleaned,
rinsed, and
thoroughly
dried
before loading into the sterilizer. Carefully
inspect all instruments and devices for cleanliness and dryness and for
flaws
or
damage prior to packaging. Ifvisible soil or moisture
is
present,
the item must be re-cleaned and dried prior to sterilization. Devices and
instruments with flaws or damage should be replaced
or
repaired before
using.
The process
of
cleaning
is
necessary to remove oTganic and inorganic soil
and debris from equipment. In this process, many microorganisms are
removed from the surface
of
the items. The process
of
sterilization
inactivates all remaining spores and live microorganisms.
Remove all blood, tissue, and soil from items
by
following the device
manufacturer's instructions using an appropriate detergent
or
cleanser.
Rinse items thoroughly to remove detergent or cleanser residue.
Dry
all items thoroughly. An acceptable method for drying is to blow
compressed gas through the lumen until
no
moisture exits the distal end
of
the device. Please ensure that any method used to dry the devices
is
in
accordance with the manufacturer's instructions for use or contact
the device manufacturer to obtain appropriate and safe procedures.
It
is
necessary to remove moisture from all parts
of
the items.
Only
dry
items
should
be
loaded into
the
sterilization
chamber
to
prevent
cycle cancellation.
28
STERRAOil!>
NX
~
User's Guide