6 Access Levels and Supervisor Tasks
Vacuum Units
torr/mtorr
expresses vacuum measurements
in
torr and millitorr.
kPalPa expresses vacuum measurements
in
kiiopascaJs and Pascals.
is
the factory default setting.
Load Data
Entry
Option
Enabled causes the Enter Load Item Data screen
to
be
displayed after
login. This
is
the factory default setting.
Disabled Enter Item Data screen.
Load Removal Option
With Login that a user enter a user identification and password to
open the sterilizer door when a cycle is complete.
Without Login allows any person to open the sterilizer door when a
is is
the factory default
Notepad Option
Enabled causes the Notes screen to be displayed after login. This
is
the factory setting.
Disabled skips the Cycle Notes screen.
Network Option
Enabled allows the sterilizer to be connected to a network.
Disabled disables the network connection. This
is
the
setting.
Alarm Volume
Touch the slider
to
adjust the volume
of
the alarm loudspeaker. Move the
slider to the far left end to turn the alarm off. The factory default
setting
is
in the middle
of
the scale.
NX'" User's Guide
54