Message
3.3 VOL T SUPPLY LOW
5 VOLT SUPPLY HI
5 VOL T SUPPLY LOW
I
AIR PUMP SENSOR
READ OFF
i AIR PUMP SENSOR
READ ON
ALARMS NOT HEARD
. A TM SWITCH STUCK
i AT ATMOSPHERE
ATM SWITCH STUCK
AT VACUUM
BAD MONITOR
BARCODE FAILURE
CANNOT TURN
PLASMA OFF
CARRIAGE SENSOR
FAILURE
NX
'"
Users
Guide
Probable
3.3 volt DC supply < 2.64
volts.
upply>
6 volts.
5 volt DC supply
< 4 volts.
Air pump mechanical failure
or air pump sensor failure.
Air pump mechanical failure
or
air pump sensor failure.
Alarms not functioning.
Atmospheric switch failure .
Atmospheric switch failure.
UV monitor not functional.
Barcode scanner did not read
barcode label.
Plasma power out
of
specification.
sensor not
functional.
Troubleshooting
UQ(lestea Remedy
Representative.
Call your ASP Customer Care
Representative.
Call your ASP Customer Care
Representati v
e.
Call your ASP Customer Care
Call your ASP Customer Care
Representative.
Verify pathway between UV
lamp and monitor is blocked.
If
the problem call your
ASP Customer Care
Representative.
Verify cassette is inserted
correctly. Verify barcode.
If
the
problem persists, Call your ASP
Customer Care
Call your ASP Customer Care
Representative.
Verify cassette
is
inserted all the
way in.
If
problem call
your ASP Customer Care
95