38
Min Lmt: A minimum current or resistance threshold that when not exceeded triggers a
failure.
Ramp Up: A length of time that is allowed for the test voltage to climb from 0 to the voltage
set point.
Dwell: A length of time that is allowed for the set point voltage to be applied.
Delay: A length of time that the set point voltage is applied but no judgment of the set
parameters is made. A judgment of the parameters is not made until the end of the delay time.
Ramp DN: A length of time that is allowed for the test voltage to decay from set point to 0.
Arc Sense: During hipot testing some low current arcing may be allowable. Arc sense is a
maximum allowable threshold for arcing.
The numbers corresponding to the different arc sensitivity levels are 0 through 9, 1 meaning
the maximum threshold of allowable arcing, 9 meaning the minimum threshold of allowable
arcing and 0=OFF. After the “+” or “-” is pressed, the arc sensitivity will automatically adjust
to the new setting.
Frequency: This parameter is available in AC tests only and is selectable using a softkey
between 50 and 60Hz.
Continuity: This function checks for a connection between the cont. check and return lead.
This is a basic DC continuity check that measures the continuity value but does not display it.
Continuity may be selected ON or “OFFusing a softkey.
Connect: This function when turned ON will connect or link the current step to the next step.
The third step in a Memory will not have this function because it is the last step of the Memory
location.