Chapter 2 - Introduction
C
ooling Specifications
Cryobar Cooling (Peltier Element) Max. - 55°C (-67°F)
Chamber Temperature Max. -35°C -2K/+1K (-31°F -2K/+1K) at +20°C room Temp.
(decrease 0,7°C/1°C room temp. increase)
Defrosting Options Programmed / Immediate
Microtome and Sectioning Specifications
mµ 005 --- 1
mµ 001 --- 1
01 --- 1 µm in 1 µm steps
10 µm --- 20 µm in 2 µm steps
20 µm --- 50 µm in 5 µm steps
50 µm --- 100 µm in 10 µm step
mµ 005 --- 5
03 --- 5 µm in 5 µm steps
30 µm --- 100 µm in 10 µm steps
02 ni mµ 002 --- mµ 001 µm steps
05 ni mµ 005 --- mµ 002 µm steps
Specimen Retraction (Return Travel) 20 µm
Vertical Specimen Stroke 64 mm
Horizontal Specimen Movement 28 mm
Specimen Approach Automated via arrow buttons
Chuck Size 30, 40 mm (additional special sizes are available)
Specimen Orientation x - and y - axes universal 8° with zero positioning; z - axis up to 360°
Coarse Feed Motorized
Cutting Drive Manual via handwheel
Handwheel Lock Mechanical
Clearance Angle Adjustment Disposable blade carrier 8- 16°