78 E1070 Service Manual
3 Using the Menus
8. Toggle between the choices by pressing the TARE or UNITS key. When your
choice is displayed, press the ENTER key…
EnAbLE is displayed.
9. Press the UNITS key…
WArn is displayed. Use this to set the zero drift as a percentage of total
capacity which will cause a warning to be logged.
10. Press the PRINT key…
Numeric entry screen appears.
11. Enter a number, in percentage of total capacity, and press ENTER to accept
it…
WArn is displayed.
12. Press the UNITS key…
Error is displayed. Use this to set the zero drift as a percentage of total
capacity which will cause an error to be logged.
13. Press the PRINT key…
Numeric entry screen appears.
14. Key in a number, in percentage of total capacity, and press ENTER to accept
it…
Error is displayed.
15. Press the SELECT key…
Z.driFt is displayed.
16. Press the UNITS key…
Err.Log is displayed. Use this to view error logs for the SensorComm
system.
17. Press the PRINT key…
oLoAd is displayed. This is the overload counter.
18. Toggle between oLoAd (overload) and uLoAd (underload) by using the TARE
or UNITS key and press ENTER when your choice is displayed…
Under both oLoAd and uLoAd are choices to view the counter (Cntr) or
print (PRINT) the report.
19. Toggle between Cntr and Print by using the TARE or UNITS key and press
ENTER when your choice is displayed…
If you chose Cntr, the overload or underload count will be displayed.
If you chose Print, you can choose the port to print through and press
ENTER. The counter value for overload or underload will be printed.
20. When you are done with the error log section, press the
SELECT key
repeatedly until…
Err.Log is displayed.