23
Model E1070 Indicator Service Manual
FILTER
(Noise filtering)
Use this item to set the noise filtering parameters.
1. From previous step 3, press the UNITS key. . .
FILTER is displayed.
2. Press the PRINT key. . .
Current setting is displayed. Choices are OFF, FLTR 1 and FLTR
2.
Off means no filtering. FLTR 1 filtering is slower response to
weight in a longer time period with improved accuracy. FLTR 2
filtering is faster response to weight in a short time.
3. Scroll through the choices by using the TARE or UNITS key and press
the ENTER key to accept the displayed choice. . .
If you choose OFF, display returns to FILTER. You can continue
to the next menu item (d.Point).
If you choose FLTR 1 or 2, continue to step 4.
4. With FLTR 1 or FLTR 2 displayed, press the PRINT key. . .
CONST is displayed. This stands for Constant and is one of two
filtering parameters you need to set.
5. Press the PRINT key. . .
Current value is displayed. For the Constant value you can pick a
value between 1 and 10. Set the number low for small vibration
problems and higher for more dampening effect.
6. Scroll through the choices by using the TARE or UNITS key and press
the ENTER key to accept the displayed choice. . .
CONST is displayed.
Service Menu—SCALE submenu (continued)