24
Model E1070 Indicator Service Manual
D.POINT
(Decimal point)
7. Press the UNITS key. . .
THRESH is displayed. This stands for Threshold, the 2nd filtering
parameter.
Threshold causes the indicator to respond quickly to large weight
changes. Threshold is the amount of weight change, in calibration
units, beyond which the filtering will be temporarily disabled. For
example, if you set this to 10 lbs, a weight change over 10 pounds
occurring during the sample time will disable the filtering until the
weight change during the sample time drops below 10 lbs.
8. Press the PRINT key. . .
Current value is displayed.
9. Key in a value. Press the ENTER key. . .
THRESH is displayed.
10. Press the SELECT key. . .
FLTR 1 or FLTR 2 is displayed.
11. Press the SELECT key. . .
BUSY is displayed briefly then FILTER. Whichever filter you set
up becomes the active filter for the indicator.
Use this item to toggle between decimal point and a comma for the fraction
delimiter for the display. For example, if you pick DEC the display will show
10.5. If you pick COMMA, the display will show 10,5.
1. From previous step 11, press the UNITS key. . .
D.POINT is displayed.
2. Press the PRINT key. . .
The current setting is displayed.
3. Toggle between the choices, DEC or COMMA, by using the TARE or
UNITS key and press the ENTER key to accept the choice. . .
D.POINT is displayed.
A THRESHOLD setting of 0 will
turn filtering on all the time.
Service Menu—SCALE submenu (continued)
Example:
decimal = 000.00
comma = 000,00