<LF>N<CR> 1st N sends the scale type; S or C.
TYP:S = scale, TYP:C = Classifier
2nd N sends CAP:uuu:c..c:n:d where
uuu = unit of measure
c..c = full capacity of this range. If multi-
interval is not enabled, this is the scale
capacity.
n = Least significant count-by digit
d = decimal point position:
0 = none
1 = XXXX.X
2 = XXX.XX
3 = XX.XXX. . .etc.
3rd N sends the following if multi-interval is
enabled: the same info as the 2nd N
except for the upper range of the multi-
interval.
3rd N sends the following if the multi-
interval is disabled: List of the SMA level 2
commands that are implemented:
CMD:PTMCU
Last valid N sends END
Subsequent N commands will return a ‘?’
response.
<LF>ESC<CR> This reboots the indicator
<LF>XP<num><CR> Request a print format to be printed
Service Menu—SERIAL submenu (continued)
The I and N commands are
related. An I command must be
sent before the first N com-
mand is sent. Multiple N
commands can be sent after
the I command and each one
will return a different piece of
data. If an N command returns
a ‘?’ or END response, an I
command is needed to reset
the N command