97
Model E1070 Indicator Service Manual
9. Press the UNITS key. . .
LOGS is displayed. The Logs menu item allows you to view, print
or clear logs for the following:
Calibration
Overload and Underload Errors
SensorComm Errors
Ghost Errors
10. Press the PRINT key. . .
CAL is displayed. You must print out this log to view it. Below is a
sample of a calibration log printout:
11. Press the F1 key to print the report. . .
Port 1 or Port 2 is displayed. Use this to choose which port to print
through.
12. Toggle between the port choices using the TARE or UNITS key and
press ENTER when your choice is displayed. . .
Report is printed and display returns to CAL.
13. Press the UNITS key. . .
ERR.LOG is displayed.
14. Press the PRINT key. . .
OLOAD is displayed. This is the overload counter.
15. Toggle between OLOAD and ULOAD (overload) by using the TARE or
UNITS key and press ENTER when your choice is displayed. . .
Under both OLOAD and ULOAD are choices to view the counter
(CNTR) or print (PRINT) the report.
LOGS
(Log submenu)
Calibration Log:
Time: 23:58
Date: 12/17/2003
Scale Serial#: 123456
Calibration Zero:
0.3456 mV/V
Calibration Span:
2.3455 mV/V @ 3000 lb
Supervisor Menu—(continued)