98
Model E1070 Indicator Service Manual
Ghost Error Log:
Cell#1 @ 14:00 on 12-17-03
16. Toggle between CNTR and PRINT by using the TARE or UNITS key
and press ENTER when your choice is displayed. . .
If you chose CNTR, the overload or underload count will be
displayed.
If you chose PRINT, you can choose the port to print through and
press ENTER. The counter value for overload or underload will be
printed.
17. When you are done with the error log section, press the SELECT key
repeatedly until. . .
ERR.LOG is displayed.
18. Press the UNITS key. . .
S-COM is displayed. This is the SensorComm error log. See a
sample below and the note at left.
19. Press the PRINT key. . .
PRINT is displayed.
20. Press the PRINT key to print the report.
Port 1 or PORT 2 is displayed. Use this item to select which port
to use for printing the report.
21. Toggle between PORT 1 and PORT 2 using the TARE or UNITS key.
Press the ENTER key when the choice you want is displayed. . .
Display shows BUSY briefly then returns to PRINT.
22. Press the UNITS key. . .
CLEAR is displayed. Use this to clear the log from memory.
23. Press PRINT to clear the log. . .
SURE? is displayed.
24. Press ENTER to clear the log. . .
CLEAR is displayed.
25. Press the SELECT key. .
S-COM is displayed.
26. Press the UNITS key. . .
GHOST is displayed. This is the Ghost error log. See a sample
below.
SensorComm Log:
Error #1: 15:00 on 12-
17-03
Misc:
Overloads:
12 overloads
Last overload:
12:00 on 12-17-03
Supervisor Menu—(continued)
See the section SensorComm
Errors to see the list of error
codes.
The S-COM log only appears if
the scale is configured as a
SensorComm scale. The Ghost
log will only appear if ghosting
is enabled.