Programming Page 14/44
MID_M2000_BA_02_1004
Quick Setup
Totalizer Unit
This parameter establishes the units of measure for the totalizers:
Full Scale Flow
This parameter sets the maximum flow the system is expected to
measure. This parameter has influence on other system parameters.
These parameters include: Frequency output and current output.
In terms of flow velocity, the meter’s limit are from 0.1 to 12 m/sec.
Moreover the values for Low Flow Cut-off and limits monitoring depend
on Full Scale Flow.
The full scale flow is valid for both flow directions.
Note: If the flow rate exceeds the full scale setting, an error message
indicates that the configured full scale range has been exceeded.
Low Flow Cut-off
Low Flow Cut-off defines the threshold at which flow measurement will
be forced to zero. The cutoff value can be from 0 % to 9.9 % of the full
scale flow. Increasing the threshold will help prevent false reading
during “no flow” conditions possible caused by vibrations or liquid
fluctuations.
Unit Unit
L Liters UKG Imperial Gallons
HL HectoLiters Lb Pounds
Cubic Meters Oz Fluid Ounces
CFt Cubic Feet Aft Acre Feet
USG U.S. Gallons BBL Barrel
MG MegaGallons

Other manuals for Badger Meter ModMAG M2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Badger Meter ModMAG M2000 and is the answer not in the manual?

Badger Meter ModMAG M2000 Specifications

General IconGeneral
BrandBadger Meter
ModelModMAG M2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals