Programming Page 22/51
MID_M2000_BA_02_1807
+
-
5.2.3 Inputs and outputs
Digital Input
Digital input lets you reset totalizers, batches, interrupt flow
measurement (Positive Zero Return) or set to ADE.
Input switching (by a normally open contact) is provided by
applying an external potential of 5 to 30 VDC or by an
internal voltage source of 24 VDC via output #2.
By using the internal source, set the function of digital
output #2 to “24 VDC Supply“. Jumper JP2 must be
placed.
Reset Totalizer
Totalizer T2 will be resetted (only if the programmed flow
direction is in uni-directional mode)
Batch Reset
Starts the dosing process.
Positive Zero Return
Stops measurement until the contact is closed (for
example during a cleaning process).
ADE
Absolute Digital Encoder (ADE
®
). Remote meter reading
technology using ASCII communications protocols. See
also menu Advanced / Encoder protocol.
8
9
5-30 V
DC
3
8