7-10
920512-768 Operator Manual Troubleshooting
6. Alarm sounds and screen displays:
ALARM: LOAD DOOR
DID NOT UNLOCK
Printer records:
*ALARM
LOAD DOOR
DID NOT UNLOCK
HH:MM:SS YY/MM/DD
or (if double door units)...
ALARM: UNLOAD DOOR
DID NOT UNLOCK
Printer records:
*ALARM
UNLOAD DOOR
DID NOT UNLOCK
HH:MM:SS YY/MM/DD
Load or unload door remain locked after DOOR OPEN
touch pad was pressed.
1. Press Silence Buzzer or ACK touch pad to silence
alarm buzzer.
2. Air supply too low – Open air supply valves.
3. Defective door safety switch – Call STERIS.*
4. Load/Unload door lock pneumatic cylinder defective,
verify operation – Call STERIS.*
5. If condition reoccurs, call STERIS.*
7. Alarm sounds and screen displays:
ALARM: LOAD DOOR
DID NOT LOCK
Printer records:
*ALARM
LOAD DOOR
DID NOT LOCK
HH:MM:SS YY/MM/DD
or (if double door units)...
ALARM: UNLOAD DOOR
DID NOT LOCK
Printer records:
*ALARM:
UNLOAD DOOR
DID NOT LOCK
HH:MM:SS YY/MM/DD
Load or unload door remain unlocked after DOOR CLOSE
touch pad was pressed or more than 5 seconds were
required to lock door.
1. Press Silence Buzzer or ACK touch pad to silence
alarm buzzer.
2. Air supply too low – Open air supply valves.
3. Defective door safety switch – Call STERIS.*
4. Door obstructed – Clear door obstruction.
5. Close load or unload door sensor misadjusted – Call
STERIS.*
6. Load/Unload door lock pneumatic cylinder defective,
verify operation – Call STERIS.*
7. If condition reoccurs, call STERIS.*
* Service charges may be incurred. Consult your warranty for details.
Table 7-4. Troubleshooting Chart - Alarm and/or Printout (Cont’d)
ALARM MESSAGE POSSIBLE CAUSE AND CORRECTION

Table of Contents