TROUBLESHOOTING
Code
Machine Screen
Guidance
Resolution
208
Incorrect Dialysate
Composition
Check set values
and connected
concentrates.
Ensure the acid jug container is filled or that the wall connection is secure and that the
right concentrate is connected to the machine.
Check for air in the concentrate line. Verify that the red concentrate connector is free
from build-up and that the blue O-rings are intact.
If needed, remove BiCart cartridge and close the latches for at least 2 seconds. If
clumping is present in the cartridge shake to eliminate.
Press the fluid key and go to the Cond tab. If the” Actual” conductivity number is not
fluctuating after a few minutes, send the machine into a Rinse. Select Disinfect key,
then select Rinse tab, then select Rinse. When completed – reconnect the concentrates
and monitor the conductivity.
If air is present in the acid line, wait – the air may need to be cleared from the acid line
by the machine. The machine will not airlock.
Note:
Pressure may build up in the BiCart cartridge. Lift the top latch to release the
pressure, reattach.
211
566
Conductivity
Out of Limits
When the dialysate
conductivity is outside
the set alarm limit.
Incorrect Conductivity
When the conductivity
is not correct during a
functional check.
Ensure the concentrate containers/BiCart cartridge are correctly connected, filled and
not empty.
Ensure there is no kinking in the concentrate line and that wands are not sucking air.
Ensure the BiCart cartridge has primed if the cartridge was dry when added to the
machine – if not, consider removing, shaking to eliminate clumping, and reattaching
to the arms.
Check acid connector and BiCart arms for buildup and/or leaking.
Press fluid button and select Cond tab and see if the concentrate selected for the
treatment is the same as the acid jug connected to the machine.
Wait until set conductivity is achieved.
Note:
When priming a BiCart cartridge during therapy, the BiCart arms must be
closed for at least 2 seconds to initiate the priming process.
607
Wrong Disinfectant
Check disinfectant.
To continue press
Confirm.
Note:
Always follow the manufacturer’s instructions when reconstituting citric acid.
Ensure the correct type and concentration of disinfectant is connected to the machine –
citric acid or bleach.
Verify that the citric acid wand is below the level of the fluid in the jug and that the jug is
not empty.
If powder is found at the bottom of the citric acid jug, remove the wand and mix the jug.
Ensure BiCart holder arms are clean and closed fully.
Ensure acid and bicarbonate concentrate connnectors are clean and fully inserted in
their ports.
Try using a different citric acid jug from a machine that did not alarm during heat citric
disinfection.
Consider that crystallization of the citric acid wand may have occurred, and the line
may have to be purged with RO water.
If unable to clear the alarm, perform a heat disinfection and contact Technical Services.
538
-
543
Dialysate Line
Sensor Test
Remove and attach the
dialysate lines from
the dialysate ports to
restart the test.
May occur if five disinfection programs have been run in a row, without running function
checks in between.
Remove the dialysate lines from the dialysate ports on the machine and wait.
Ensure the lines are reseated properly.
Once reseated, the test will continue and message on screen will clear shortly.
571
Leakage Test Failed
Check dialysate lines.
To continue press
Confirm.
Appears when the dialysate lines are not properly attached to the dialysate ports.
Check the dialysate lines are properly attached to the dialysate ports.
Confirm the attention.