OTHER FUNCTIONS
35
34
AK 98 can be placed into concentrate standby mode. In concentrate standby mode, the concentrate
is no longer used, and water consumption can be reduced. Concentrate standby mode can be
activated when the functional check is done and the flow diagram turns green. When concentrate
standby mode is active, a message will appear in the information field. The bypass path of the flow
diagram will turn orange. After 1 hour in concentrate standby mode, the machine will automatically
start preparing dialysate again. If blood is detected, concentrate standby mode cannot be activated.
Concentrate standby mode can be activated manually, or preset to activate automatically.
Recirculation and concentrate standby mode cannot occur simultaneously.
How to Activate / Reactivate Concentrate Standby Mode Manually
1. Press the Fluid button.
Note: The machine can be preset to automatically activate
concentrate standby mode.
2. Select the Dialysate tab.
3. Activate concentrate standby mode.
4. If reactivating, repeat every hour as needed and allowed by
your procedure.
How to Exit Concentrate Standby Mode
1. Select the Information tab.
2. Press Confirm the message.
Note: Do not stop the blood pump to Connect patient until
the dialysate line is green.
Concentrate Standby Mode (CSBM)