HOW TO USE
• The main power cable must be connected to the main power supply.
• The main switch must be turned on.
• The water supply must be connected to the inlet water and powered on.
• The outlet tube (drain) should be properly placed with an air gap between the dialysis machine and the drain.
• The dialysate lines should be connected to the dialysate ports.
• Acid concentrate, BiCart cartridge, dialyzer, bloodline, and saline must be ready for use.
3. Before connecting concentrates, wait for the machine prompt to
connect concentrates.
4. Once prompted to connect concentrates, remove the red concentrate
connector from the dialysate port.
5. Attach the white wand to the red concentrate connector. Place the wand
in the acid concentrate container.
6. Select the concentrate combination to be used from the preset options
on the screen.
Note: The user should ensure a current listing (brand and
composition) of specified concentrates in the dialysis machine
is the same as actually used during treatment, and that it is
the same as is prescribed for the patient.
Note: Incorrect choice of dialysate concentrate may cause incorrect
composition of the dialysate. Incorrect composition may lead
to electrolytic imbalance in the patient's blood.
7. Press the release buttons to release the latches. Open the BiCart cartridge
holder. Pull the upper latch out and up. Pull the lower latch down.
8. Remove the caps from the BiCart cartridge and attach it narrow end down.
9. While supporting the lower latch, press down the upper latch to secure the
cartridge.
10. Attach the dialyzer in the holder. If necessary, squeeze the spring-
clip to place the dialyzer in the holder. Move the dialyzer caps to
the dialysate ports on the dialyzer to prevent saline loss. Follow the
dialyzer's instructions for use.
11. Hang the saline bag on the IV pole.
1. Gently press the On/Off button to turn on the machine.
2. Before a treatment can be started the dialysis machine performs a
functional check. Do NOT attach blood lines or open the blood pump
door until the blood pump button is flashing.
Pre-treatment Check
Functional Check
7
Operation Start Up
WATCH VIDEO