HOW TO USE
Circuit Preparation and Line Connection - Attaching the Arterial Blood Line
1. Open the blood pump door. Press the center of the
cover and pull the tab.
2. Place the blood pump segment into the blood pump. Do NOT
push/thread the segment into the blood pump rotors. Ensure the
arterial pressure transducer points down and the heparin line is
at the top.
6. Remove the protective caps and attach the end of the arterial
blood line to the dialyzer. Ensure the arterial blood line is securely
attached to the dialyzer.
Note: Ensure the arterial blood line is not cross threaded and
that it's firmly attached to the dialyzer to prevent patient
blood loss.
7. Route the line around the line guide.
8. Insert the arterial blood line into the arterial clamp (marked with a
red dot).
9. Gently pull the blood line down to secure in place.
Note: Ensure the blood line between the arterial clamp and blood
pump is not too stretched.
3. The blood pump segment collars must be outside of the pump
housing.
4. Hold the segment in place and close the blood pump door.
5. Follow the arterial guide arrows and gently press the blood line into
the upper guide.
8