Chapter 5: Troubleshooting 103
Tube stuck on SIT,
software not responding
• Loader not set up for
automatic loading
• Software problem
Manually remove the tube from
the SIT.
1 Hold the tube and move the
aspirator arm aside.
2 Gently pull the tube off the SIT
and replace in carousel.
3 Release the aspirator arm.
4 Restart the worklist run.
5 If needed, modify the cytometer
for automatic loading. See
Changing to Automatic
Loading on page 41.
Carousel not rotating
correctly
Carousel not engaged with
alignment pin on drawer
Rotate the carousel on the spindle
until the alignment guide pin
engages with the alignment hole,
and press down. See step 4 on
page 38. If the problem persists,
contact BD Biosciences for
assistance.
Aspirator arm bar vertical Rotate the bar horizontally. See
Changing to Automatic Loading
on page 41.
Loader Troubleshooting (continued)
Observation Possible Causes Recommended Solutions