Chapter 2: BD FACS Loader Option 37
2 If needed, remove the carousel from the Loader:
• Open the Loader cover.
• Pull out the Loader drawer.
• Place your thumb on the center spindle, and press down with your
thumb while gently pulling up on the carousel handle with two fingers,
as shown.
3 Vortex the sample tubes and place them uncapped in the carousel(s)
according to the worklist.
For accurate results, match the tubes to those listed on the printed worklist
(BD FACSCanto clinical software) or the Carousel Assignment tab
(BD FACSDiva software). Print a copy of tube assignments and use the
printout as a guide when filling each carousel.
To prevent binding during loading, make sure label thickness per tube is
less than three labels. Flatten labels completely before placing a tube in the
carousel.