Chapter 4: Maintenance 71
Priming Fluidics Lines
Use the Prime After Tank Refill command to remove air from the fluid lines after
you change a cubitainer or detach the fluidics lines for other maintenance.
1 Select Cytometer > Cleaning Modes > Prime After Tank Refill.
2 Select the checkboxes for the cubitainers you changed; click OK.
3 Click OK when the completion message displays.