Chapter 4: Maintenance 79
2 A dialog indicates Bubble Filter Purge progress. Wait until the purge
finishes.
3 Click OK when the completion message is displayed.
4 Repeat steps 1 through 3 until bubble-free liquid enters the flow cell from
the bubble filter.
5 When finished, select Cytometer > Cleaning Modes > De-gas Flow Cell.
BD FACSCanto clinical software
BD FACSDiva software
BD FACSCanto clinical software
BD FACSDiva software