Chapter 4: Maintenance 89
Procedure
1 Ensure that the cytometer is not acquiring events.
2 Detach the sensor and fluid line from the cart.
3 Unscrew the cap on the cubitainer.
4 Remove the level sensor assembly and discard into a suitable receptacle.