EasyManua.ls Logo

Bedfont NoxBox - Page 34

Bedfont NoxBox
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34
16. Ifthesystemtestfailed,followthediagnostictipstocorrect:
a. Ensure that all connections on the Test Line Assembly are made as directed.
b. EnsuretheNOxFlowsensorisorientatedwiththeairowinthedirectionofthe
arrowmarking,thedoseportionclosesttothesampleendofthetestline.
c. EnsurethevalvesareopenonallNOsupplycylinders.
d. Checktheairowissetto10L/min.
17. Ifthesystemtestfailsasecondtime,theunitcannotbeusedforintelligentautomated
delivery.CalltheSystemEngineer,andreplacethesystem.
18. Oncethesystemtesthaspassed;shutofftheO
2
ow,anddisconnectthetestedLine
Assemblyfromthesupplyline.TheveriedLineAssemblyisreadytoconnectintothe
patientventilatorcircuit
If the NOxBOX
i
system is not to be used within next 10 minutes, turn
off the cylinder valves, detach the supply lines from Port 1 and Port
2, and release the line pressure using the purge needle at the back of
the monitor. Then re-attach the supply lines to Port 1 and Port 2.
WARNING: If the NOxBOX
i
system is left pressurized for more than 10
minutes while not in use, repeat the system test. Connect the Line
Assembly to the air supply as described in steps 11-17. Press Next.
The system test will run automatically.
Ventilator inspiratory limb showing NOxFlow placement before humidier.
19. Connect the NOxBOX
i
tothepatientventilatorinspiratorylimb.SeeSection5.3onwardsfor
Patient Circuit connection diagrams.
The sample point adaptor should be connected immediately behind the patient Y-piece on
theinspiratorylimb.Toachieveconsistentresults,placeaone-wayuttervalvebetweenthe
sampleconnectorandtheY-piecetopreventcontaminationofthesamplefromexhaled
breath.
20 . Connect the upstream end of the NOxFlow sensor to the upstream inspiratory limb tubing
TheNOxFlowshouldbesituatedupstreamofthehumidier.
Maintaintheoptimum1mdistanceofventilatortubebetweenthesamplepointandthedose
introductionpointontheNOxFlowforbestperformance(between0.7and1.3misideal).
The system is now ready for dose entry and alarm setting procedures.

Related product manuals