79
Critical
Delivery
Fault
The NOxBOX
i
has detected a critical
faultwithintheintelligentdelivery
system,andcannolongerguarantee
safedeliveryfunction.
Checktheventilatorisconnectedand
supplyingsufcientow.
Ifanothersystemisnotreadilyavailableand
patientrequirestherapy,engagethe
manualoverridemode,replacethesystemas
soonaspracticallypossibleandalerttheService
Engineer.
Touch
screen will
not
respond.
The NOxBOX
i
has detected a
critical fault within the intelligent
deliverysystem,andcannolonger
guaranteesafedeliveryfunction.NO
delivery to the patient may have
stopped.
Ifanothersystemisnotreadilyavailableand
patientrequirestherapy,engagethemanual
overridemode,replacethesystemassoonas
practicallypossibleandalerttheService
Engineer.
Medium Priority Alarms during therapy
Alarm Priority Possible Cause Recommended Action
Zero
Calibration
Every24hoursduringusetheNOxBOX
i
prompts the user to perform a sensor
zero. This operation ensures the most
accurate system performance by
checking the gas sensor reading
performance.
NOTE:Thistesttakesupto2minutesto
perform. During this time the monitored
patientgaseswillbeofine.The
NOxBOX
i
continuestodeliverNOduring
this time.
Nospecialconnectionsarerequired;thezero
calibration is fully automatic.
Press the tick to start the zero calibration.
The zero calibration can be delayed if the
system is not currently in a stable dose
deliverystate:e.g.Ifthepatientdosehas
recently been changed and the system is still
stabilisingtothenewdoselevel,dismissthisalarm
message and perform the zero when the notice
next appears.
Zero
Calibration
Fail
Oneorallofthesensorshavefailedthe
zero(low)calibration.
Ambient conditions may be affecting the
zero sample.
CheckambientNOxAIRmonitorforhighlevels
ofNO.Ifhighlevelsaredetected,checkthe
regulator(s)andsupplyline(s)forleaks.
Asensormayhavebecomeunstable
or residual gas may be present in the
system.
Repeat zero calibration.
Checkthezeroport(rear)hasnotbeen
blocked.
Ifanothersystemisnotreadilyavailableand
patientrequirestherapy,engagethemanual
overridemode,replacethesystemassoon
aspracticallypossibleandalerttheService
Engineer.
Cylinder
Low
This will appear when an alternate
viablegassupplyisdetected,butthe
current feed cylinder is nearly depleted.
Replace the cylinder with a fresh
supplytoresolvethisalarm.
Alternatively,oncethecylinderisempty,close
thecylindervalvefully,removethefeedhose
from the rear of the system and release the
pressure using the purge needle on the
monitor.Pleasenote,oncethesecond
cylinderbeginstodeplete,ifthisrstcylinder
hasnotbeenreplacedintheinterimperiod,
the‘cylindersupplycritical’alarmwillbe
triggered.