8.3 Alarm Priority- Alarms during therapy
Thebelowareaseriesofalarmswhichmaybeseenduringtherapy(oncethedevicehasbeen
setup)ifanissueisdetectedwiththeNOxBOX
i
.
Alarm Priority Possible Cause Recommended Action
NO Low MonitoredlevelsofNOgas
beingdeliveredtothepatienthave
dropped below the alarm setting
boundary.NOxBOXideliverysystem
cannot mainatin correct dose setting.
Checksamplelineiscorrectlyattachedtoventilator
circuit and NOxBOX
i
water trap inlet.
Check Sample line for blockages.
Checkwatertrap(includingbarrelthread)for
damage and/or leaks.
Checknoventilatorcircuitbreaksorleakageshave
occurred.
Checksupplycylinderisconnected,open;thereare
no leaks and the concentration matches the system
setting.
Check correct orientation of NOxFLOW.
Check NOxFLOW dose line and
connection is connected and there are no
blockages or leaks.
CheckNOxFLOWowdetectionlinesand
connection(includingOrings)areconnectedand
there are no blockages or leaks.
Theventilatorminutevolumemaybe
too low.
Checkventilatorminutevolume(seesection3.2for
owranges),youmayneedtoincreasethe
ventilatorbiasow.
The NO low alarm may be
inappropriately set by user.
CheckNOlowalarmvalueandreducevalueif
ventilatorsettingsdeemnecessary.
The NO sensor may require
replacing.
Ifanothersystemisnotreadilyavailableandpatient
requirestherapy,engagethemanualoverride
mode,replacethesystemassoonaspractically
possibleandalerttheServiceEngineer.
NO High MonitoredlevelsofNOgasbeing
deliveredtothepatienthaverisen
abovethealarmsettingboundary.
NOxBOX
i
deliverysystemcannot
maintain the correct dose setting.
Check supply cylinder concentration matched the
systemsettings.Ifpossible,changetheNOsupply
cylinderforthecorrectconcentration.Ifnot,call
ServiceEngineertoresolve.
Checknoventilatorcircuitbreak/leakagehas
occured that may cause build-up of NO
concentrationduetolackofventilatorow.
Check correct orientation of NOxFLOW
CheckNOxFLOWconnection(andOrings)to
NOxBOX
i
.
The NO high alarm may be
inappropriately set by user.
CheckNOhighalarmvalueandincreasevalueif
ventilatorsettingsdeemnecessary.
The NO sensor may require
replacing.
Ifanothersystemisnotreadilyavailableandpatient
requirestherapy,engagethemanualoverride
mode,replacethesystemassoonaspractically
possibleandalerttheServiceEngineer.
NO
2
High
MonitoredlevelsofNOgasbeing
deliveredtothepatienthaverisen
abovethealarmsettingboundary.
HighNOdosesettingsonlowventilatorowswith
High O
2
content may cause higher NO
2
build-up
thanexpected.Increaseventilatorbiasowtohelp
reducestagnationindelivery.