77
Poor quality NO cylinders can
containhighlevelsofNO
2
Connect a second supply cylinder to the alternate inlet
port.Openthecylinderanddisconnecttheprevious
cylinder,forcingacylinderchangeovertoseeifthis
resolvestheissue.
The NO
2
high alarm is set to a
defaultvalueof1.0ppm. NO
2
is
extremely toxic and poses risk to
patient health,
Thealarmvaluecanbeincreasedtoamaximumof
5.0ppm if required. Please see INO guidelines for
more information on maximum NO
2
valuesduringINO
therapy.
Incorrect placement of
NOxFLOW and sample line.
Seeventilatorcircuitdiagramforcorrectplacementof
NOxFLOW and sample line.
Whilst in standby mode NO
2
can
build up in supply lines.
Purgesupplylines(seecylinderchangeprocedure).
Stagnant gas in manual bag circuit
causing NO
2
.
Purge manual bag circuit before connecting to
patient(seemanualbaggingprocedure).
The NO
2
sensor may require
replacing.
Ifanothersystemisnotreadilyavailableandpatient
requirestherapy,engagethemanualoverridemode,
replace the system as soon as practically
possibleandalerttheServiceEngineer.
O
2
Low MonitoredlevelsofO
2
gas being
deliveredtothepatienthavefallen
below the alarm setting boundary.
Check sample line for blockages.
Checkwatertrap(includingbarrelthread)for
damage and/or leaks.
Checksamplelineiscorrectlyattachedtoventilator
circuit and NOxBOX
i
water trap inlet.
Checknoventilatorcircuitbreaksorleakageshave
occurred.
TheNOgasisbalancedinN2,this
is an asphyxiant gas. At high NO
doselevelsforlowconcentration
cylinders(e.g.200ppm)thelevel
ofgasdeliveredintotheventilator
streamcanreducethe%v/vofO
2
beingdeliveredtothepatient.
Check O
2
concentrationsettingatventilator.
AdjustO
2
alarmvalueifdeemednecessary.
The O
2
sensor may require
replacing.
Ifanothersystemisnotreadilyavailableandpatient
requirestherapy,engagethemanualoverridemode,
replace the system as soon as practically
possibleandalerttheServiceEngineer.
Water
trap full
Watertrapislledwith
condensate from sample line. If the
water trap is allowed to
overow,thesamplepathwillblock
and water ingress to the NOxBOX
i
system could damage the internal
mechanisms and gas sensors.
Deliveryaccuracyiscompromised
and patient safety could be put at
risk.
Use disposable male-luer lock syringe contained in
NOXKITtoemptyuidfromwatertrapviathe
self-sealing drain tap located at the bottom of the water
trap. Dispose of entire syringe and contents according
tolocaldirectives(e.g.sharpswaste).
Thewatertrapusesasmalloattoactivatethealarm,
if no moisture is present gently tap the
barreltoseeifthealarmoatisintheoffposition.