EasyManua.ls Logo

Bedfont NoxBox - Zero Calibration During Delivery

Bedfont NoxBox
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60
Sensor Bias Lost
System sensors require a constant
verylowtricklechargetomaintaintheir
calibration. The system will extend
biassettlingtime.Intheeventthatthe
system is not stored on mains power
charge,afteranextendedperiodof
time the battery may completely
discharge and the sensors will lose
their calibration bias.
Ifthishasoccurred,thesystemwill
display this screen at start up.
The system must be fully calibrated
prior to use to ensure correct and
safe system performance. Contact
theServiceEngineer.
High Dose Set
SettingtheNOdoseabove20ppm
will display warning messages so the
user is aware that a high clinical dose
setting has been selected.
Prior to administering a high dose
settingtothepatient,thesystemwill
display this warning to alert the user
to the dose setting selected.
Stop Treatment Set
To stop administering Inhaled NO
Therapy,settheNOxBOX
i
doselevel
to zero.
The system will display warning
messages so the user is aware that
thestopdeliveryoftreatmentmode
has been selected.
Priortostoppingthetreatment,the
system will display this warning to
alert the user.
6.10 Zero Calibration During Delivery
Zero Calibration is an important maintenance test that the system
can run without interruption to the NO dose delivery function. Zero
calibration takes approximately 2-3 minutes to complete, during
which time the monitoring and alarm functions will be ofine. No
special connections or tools are required to perform zero calibration,
just follow the simple on-screen instructions.
Calibration Overdue Warning
The system will display this warning
if user proceeds without calibrating
whenthecalibrationisoverdue.

Related product manuals