EasyManua.ls Logo

Bedfont NoxBox - Main Screen and Alarms

Bedfont NoxBox
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
51
Gas monitoring MeasuredvaluesofNO,NO
2
&O
2
beingdeliveredtopatient
Gas alarm setting display
Gasalarmadjustmentbuttons
Gas alarm mute buttons
Dose and status Current NO dose setting
NO dose edit button
NO supply cylinder indicators
NO supply cylinder concentration setting reminder
Treatment time clocks
Battery status icon
6.7.1 Detailed areas of main screen:
Gas monitor displays the measured
gasconcentrationsbeingdeliveredto
the patient.
Each gas has alarm settings. These
can be altered by pressing the
adjustmentbuttons.
Themutebuttonbecomesactivewhen
a gas alarm is triggered.
Thedosesettingtobedeliveredtothe
patient is displayed in the NO Dose
box.
Press the edit button to change the
delivereddose.
The NO supply cylinder status shows
cylinders connected to the NOxBOX
i
.
The supply indicators warn when the
cylinders are low. Check the pressure
gaugeontheregulatortoconrm.
The system concentration setting must
match the supply cylinders. Call the
serviceengineerifthisneeds
adjusting.
WARNING: The cylinder status is not an accurate representation of the
gas remaining and therefore the cylinder pressure gauges should be
checked to conrm.
Supply
Indicator
(Gas available)
Supply
Indicator
(depleted)
Cylinder 1
Cylinder 2
Supply concentration
Top bar Displays system time and date
monitored
value
lower alarm
adjustment
lower alarm
value
mute button (inactive)
units
upper alarm
adjustment
upper alarm
value
dose setting
edit dose

Related product manuals