NOTE 1: U
T
istheA.C.mainsvoltagepriortoapplicationofthetestlevel
NOTE 2: At80MHzand800MHz,thehigherfrequencyrangeapplies
NOTE 3: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorptionandreectionfromstructures,objectsandpeople.
TheISM(industrial,scienticandmedical)bandsbetween150kHzand80MHzare6.765MHzto6.795MHz;
13.553MHzto13.567MHz;26.957MHzto27.283MHz;and40.66MHzto40.70MHz.
ThecompliancelevelsintheISMfrequencybandsbetween150kHzand80MHzandinthefrequencyrange80
MHz to 2.5 GHz are intended to decrease the likelihood that mobile/portable communications equipment could
causeinterferenceifitisinadvertentlybroughtintopatientareas.Forthisreason,an
additional factor of 10/3 Is used in calculating the recommended separation distance for transmitters in these
frequency ranges.
Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio(cellular/cordless)telephonesandland
mobileradios,amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywith
accuracy.ToassesstheelectromagneticenvironmentduetoxedRFtransmitters,anelectromagneticsitesurvey
shouldbeconsidered.IfthemeasuredeldstrengthinthelocationinwhichtheNOxBOX
i
is used exceeds the
applicableRFcompliancelevelabove,theNOxBOX
i
shouldbeobservedtoverifynormaloperation.Ifabnormal
performanceisobserved,additionalmeasuresmaybenecessary,suchasreorientingorrelocatingtheNOxBOX
i
.
Overthefrequencyranges150kHzto80MHz,eldstrengthshouldbelessthan3V/m.
2.5 Network Connections
Thedeviceisnottobeconnectedtoanetwork.
ThesystemisprovidedwithaUSBport.Thisissolelyforthepurposeofdownloadingsystemlogs
forrecordandanalysisbyasuitablyqualiedSystemsEngineer.
2.6 Applied Parts
Thesystemhasnoappliedparts.HowevertheNOxFlowandSampleLinearetreatedasapplied
parts(apartfromlabellingrequirements)asperBSEN60601-1:2006Section4.6.
2.7 Isolation from Mains
WARNING: Operator should not contact the patient whilst simultaneously being in
contact with the NOxBOX
i
device.
Shutdownofthesystemiscoveredinsection6.13Shutdown.
In order to isolate the system from the mains unplug the Mains Power Supply. Do not position the
systemsothatthisisdifculttodo.