26
4.4 Connecting NO Gas Cylinders
1. Using the on-screen Quick
StartGuide(QSG)
instructions set up the
systemasfollows:
NOTE:forexperiencedusers,the
“SkiptoSystemTests”buttonwill
bypass the on-screen QSG and
skiptosystemsetuptests).
Alternativelythe“skiptodose”
button will bypass both on-screen
QSG and the system test.
See Chapter 6.3 for detailed
instructions for using the touch
screen interface.
2. Check that the
NOxBOX
i
NO cylinder
concentration matches
thevalueshownonthe
system screen. This is
thevalueuponwhich
deliverycalculations
will be based. Do not use
any supply cylinder that
does not match this
value.
Ifthereisaproblem,
contactyourService
Engineertoresolvebefore
attemptingtodeliverNOto
a patient.
3. Check the regulator
sealing surfaces and
connector thread. If your
model has non-metal
sealingelements,ensure
that they are present and
undamaged,showingno
cracks,pock-marks,etc.
1000