80
Manual
Override
ManualOverridemodeisengaged.
The system alarms to alert the user that
thesystemisnotdeliveringinintelligent
mode.Changestotheventilatorsetting
or patient demand cannot be
automatically corrected on the system. A
specicdosesettingcannotbedialledin
andachieved.
The patient must be closely monitored
and ensure the gas alarms are correctly
settoalertforanyabnormalgasdelivery
behaviour.
Thisalarmwillresolvewhenthesystemis
returnedtonormalIntelligentDeliveryMode.
8.4 General Troubleshooting
Issue Possible Cause Recommended action
NOxBOX
i
turns on and off
immediately. NOxBOX
i
attempts to start up but shuts
down.
NOxBOX
i
won’tturnonatall.
Low battery power Connect NOxBOX
i
to mains power and turn
on NOxBOX
i
.
Aninternalfaulthasoccurred;the
system shuts down to protect
integral components.
Ifanothersystemisnotreadilyavailable
andpatientrequirestherapy,engagethe
manualoverridemode,replacethesystem
as soon as practically possible and alert the
ServiceEngineer.
Zero Calibration Fail Oneorallofthesensorshave
failedthezero(low)calibration.
Ambient conditions may be
affecting the zero sample.
Check ambient NOxAIR monitor for high
levelsofNO.Ifhighlevelsaredetected,
checktheregulator(s)andsupplyline(s)for
leaks.
Oneorallofthesensorsmayhave
become unstable or residual gas
may be present in the system.
Repeat zero calibration.
Checkthezeroport(rear)hasnotbeen
blocked.
Ifanothersystemisnotreadilyavailable
andpatientrequirestherapy,engagethe
manualoverridemode,replacethesystem
as soon as practically possible and alert the
ServiceEngineer.
Fluctuations/ Oscillations in
excess of 3ppm.
May be due to NOxFLOW Ensure O-ring and NOxFLOW are present
and connected.
Massowsensordueforservice.
ContactServiceEngineer.
HFO Frequency
Adjustthefrequencyslightlybasedupon
yourclinicaljudgement.