13
The NOxBOX
i
isonlyintendedforusewith,andconnectionto,thespeciedaccessoriesand
single-patient use disposables as detailed within this technical guide.
3.2 Specication & Environment
NO 0-99.9 ppm
NO
2
0-20.0 ppm
O
2
0 – 99.9 %
NO Delivery Range (set point)
Measuring Range
0.6–80ppm(Notethesystemallowssettingat<=0.5
ppm,accuracyguresbelowarefor>=0.6ppm,useofa
lowersettingisatcustomerresponsibility)
NO input concentrations
200,225,250,300,400,450,500,800,900&1000ppm
NO Delivery and Measurement resolution
0.1 ppm
Means of disconnection
Mains power supply plug
Mode of Operation: Synchronous(1.5-50Ltr/m)orContinuous
Sample Flow Rate:
Approx. 225ml/min
Gas Concentration Detection Principle: Sealed electrochemical sensor
Accuracy: Monitoring:
NO&NO
2
:+/-2%or0.2ppm
whicheveristhegreaterplusthe
accuracy of the calibration gas
O
2
+/-3.5%
Automatic Mode
Delivery
(measured
dose):
At the point of sampling at the
patient Y-piece the accuracy is that of
themonitoringsystemabove.
Delivery(with
respect to
set-point):
Thesystemwillservethedelivered
concentrationtotheset-pointsubjectto
theowfromtheventilator,seesection
“MaximumNODoseDelivery”belowfor
the set point range.
Once the measured dose and set point
agree then the accuracy w.r.t. the set
point is that of the monitoring system.
Until the set-point is reached then
the accuracy is that of the difference
between set-point and measured dose
plus the accuracy of the monitoring
system.
Manual Mode
Delivery(with
respect to
valueadvised
by the Dose
Calculator)
Dosecalculator,continuousow.
Thedelivereddosewillbe+/-30%of
thegureadvisedbytheonscreendose
calculator. Monitoring can be used to
determinetheactualvalue.
Intermittent(non-continuous)ventilation.
Accuracyhasonlybeenveriedwith
acontinuousowuse.Intermittent
owaccuracydosesvaluesmustbe
determined by the customer.
Display:
Touch screen LCD
NOxBOX
i
Standby Mode
Set 0ppm