14
Alarms:
Audibleandvisible
Set-up time:
< 10 minutes
Response time:
< 10 seconds to 90% FSD NO
< 40 seconds to 90 % FSD NO
2
< 15 seconds to 90 % FSD O
2
Operating Temperature:
10
o
C – 35
o
C
Operating Humidity:
15 % - 85 % RH non-condensing
Storage Temperature:
0
o
C – 40
o
C
Storage Humidity:
15 – 90% non-condensing
Operating & Storage pressure: 800 – 1200 mBar
Sensor Operating Life: 2 years
Sensor Resolution:
NO&NO
2
-0.1ppm
O
2
-
1m Bar O
2
Battery:
4 hours operation
Li-ion Polymer
745396/3900mAh/22.2V
Not User replaceable
Power Input:
EXM 80 5121
100–240VAC,50/60Hz,1.7A
Dimensions: 330 mm x 250 mm x 108 mm
Weight: 6.2 kg
Construction:
Rigidpolyurethane(PU)
WARNING:ForhighconcentrationsofNOexceeding80ppm,exposureshouldbelimitedtoamaximum
of 4 days in order to reduce the risk of sensor zero drift due to prolonged exposure. The NOxBOX
i
protects
usersfromtheadverseeffectsofsensordriftbyperformingadailysensorzerocalibration.Ifasensorhas
drifted outside of acceptable limits the zero calibration will fail.