8 Commissioning
8.1 Mains connection
Plug the mains connection cable into a shock-proof
socket (230 V, 50 Hz, 16 A)
Set the mains switch to position "I”.
The appliance tester carries out a self-test and shows the
Registration / login" screen after the start sequence.
8.2 Login
Navigate to the "User" screen by touching the "Name" input
line.
Select the user "admin" by touching the corresponding line.
You will get back to the “Registration / login” screen.
Touch the "Password" input line and enter the
preconfigured default password.
Default password of the admin user: Benning
After entering the password, you can optionally enter an order
number.
(The order number appears in the test report later.)
Confirm your entries by touching the
Confirm entry” soft key.
The “Main menu” screen opens.
Incorrect password entry
If the password has been entered incorrectly, the following
message appears on the “Registration / login" screen:
Wrong name or password!
Correct your entry and confirm the login again.
Forgot password...
Touch the “Forgotten?” button on the
Registration / login" screen.
The “Forgot password...” screen opens.
Administrator
User
Please contact the
BENNING support/ helpdesk
calling the following hotline no.:
+49 2871 93 555
Please contact a device
administrator to ask for
the current password.
Please have the following
information ready before you
call:
- Proof of purchase of the
appliance tester
If the administrator is not
available:
Follow the steps to be
taken as administrator.
Provide the BENNING support
with the serial number, user
level and index.
The BENNING support will give you a key.
Enter the key in the Password line and confirm with the
Log on” button.
The appliance tester is enabled.
8.3 Main menu
Via the main menu, you can access all functions and settings of
the appliance tester.
8.4 Calibration
After initial commissioning, the appliance tester must be
calibrated using the internal calibration.
The ST 755/ ST 760 must be calibrated every 12 months
by the manufacturer or in certified calibration
laboratories
(see chapter 30, Maintenance / calibration).
Follow the instructions in chapter 9.6 to calibrate the appliance
tester.
8.5 Change of test station