Density Meter LB 444 SECTION 5. GETTING STARTED
44
2. Select menu Parameter / Product Data via the key <
more
>.
2.1 Select Parameter menu with <
sk1
> and with <
more
> call the parameters
sequentially.
Select product
Select detector and isotope
Measuring path in product (e.g. pipe diameter) in cm
Select Measure mode (i.e. continuous)
Enter time constant (e.g. 60 s)
Disable or enable rapid switch-over
Disable or enable Interference Radiation
Default max. countrate (e.g. 100000 I/s)
Default min. countrate (e.g. 0 I/s)
For TC via current input:
Select current input range 0-20mA or 4-20mA
Enter temperature value for 0/4mA and 20mA
Return to the menu group with <
done
>.
2.2 Select Product Data menu with <
sk2
> and with <
more
> call the parameters
sequentially
Select product
Select current output 0 –20 or 4 – 20 mA
Define current output limit values 0/4 and 20 mA
Define current output error message
Define Relay 2: Function (Min)
Switchpoint (e.g. 1.2 g/cm
3
)
Hysteresis (e.g. 5 %)
Define Relay 3: Function (Max)
Switchpoint (e.g. 1.6 g/cm³)
Hysteresis (e.g. 5 %)
Only for suspension measurements
Enter solids density in g/cm³ for the selected product
Enter liquid density in g/cm³
Activate Water TC when solution is to be measured with water as
carrier liquid.
Enter reference temperature
In all other cases:
Calculate and enter temperature coefficient TC1 (linear)
Calculate and enter temperature coefficient TC2 (square)
Enter reference temperature
Return to the menu group with <
done
>.