Density Meter LB 444 SECTION 10. APPENDIX
92
10.2 Parameter Protocol
EG&G Berthold 19.071997 / 14:44
LB 444 D E N S I T Y - M E T E R V 1.00 a
Operating mode : Without Temp. Comp.
Measure mode : continuous
Parameters
Probe type : No. 0
Isotope : Am241
Measuring path : 2.0 cm
Time constant : 10.0 sec
Rapid switch-over : OFF
Sigma (Rapid) : 3.0000
Interference radiation : OFF
Sigma (Inter.) : 5.0000
Maximum countrate : 100000 cps
Minimum countrate : 0 cps
Current input
Temp. limit values: 0 - 20 mA
0/4 mA : 0.0 deg C
20 mA : 100.0 deg C
Product selection : internal
Product data
Product : 1
Current output
Current output range : 0 - 20 mA
0/4 mA : 3.0 g/cm³
20 mA : 5.0 g/cm³
Error current : -1 mA
Relay 2 function : Minimum
Relay 2 threshold : 0.0 g/cm³
Relay 2 hysteresis: 1.0 %
Relay 3 function : Maximum
Relay 3 threshold : 3.0 g/cm³
Relay 3 Hysteresis: 1.0 %
Calibration
Unit : g/cm³
Suspension measurement : Yes
Calibration data : Countrate Density Product temp.
Cps g/cm³ deg. C
Calibration product 1 : 3000 1.0664 25.00
Calibration product 2 : 2000 1.1422 25.00
Calibration product 3 : 0 0 25.00
Calibration product 4 : 0 0 25.00
Calibration product 5 : 0 0 25.00
Calibration product 6 : 0 0 25.00
Calibration product 7 : 0 0 25.00
Calibration product 8 : 0 0 25.00
Calibration product 9 : 0 0 25.00
Calibration product 10 : 0 0 25.00
Calibrate mode : automatic
Zero countrate I0 : 1 cps
Coefficient a1 : 0.00000
Coefficient a2 : 0.00000
Coefficient a3 : 0.0000e-00
Result qf : 0.00000
Factor : 1.00000
Offset : 0.00