Density Meter LB 444 SECTION 10. APPENDIX
91
10. APPENDIX
10.1 Configuration Checklist
TAG NO.
Detector
Source
FUNCTION RANGE
FACT. SETTING
FINAL SETTING
Password 0-999999
Date YY.MM.TT actual date
Time HH.MM. actual time
Model/Version e.g.: LB 444 Version
Language German/English English
Operating Mode none/lin/squ/cub/auto lin
Isotope Co/Cs/Am/Cm/Kr/Sr Am
Time Constant (s) 0.2 to 200 20
Rapid Switch-Over OFF / ON OFF
Sigma Rapid Switch-Over 0 - 9.9999 4.0000
Interference Radiation OFF / ON OFF
Sigma Interf. Radiation 0 -10 5
Pulse Rate, max 0 - 999999 100000
Pulse Rate, min 0 - 999999 0
Current Output 0-20 mA or 4-20 mA 4 -20 mA
Current on Error Hold / Value 0-22 mA Hold
Relay 2 (Min Alarm) 0-100% 10
Relay 2 (Hysteresis) 1-10.0 % 5
Relay 3 (Max Alarm) 0-100% 90
Relay 3 (Hysteresis) 1-10.0 % 5
Calibration values
Density reading
%
Pulse rate
rel.
Value pair 1
Value pair 2
Value pair 3
Value pair 4
Value pair 5
Value pair 6
Value pair 7
Value pair 8
Value pair 9
Value pair 10