Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
3
2. SYSTEM DESCRIPTION
2.1 Use and Function
The radiometric density measuring system can be used to measure the density
of
liquids
suspensions
pulps and
bulk goods.
The measurement can made directly in-line or in a tank or container. Pressure
and viscosity fluctuations or the flow rate of the product does not affect meas-
urement accuracy.
Special instrument configurations and calculations allow the measuring system
to be tailored to the local needs and conditions of the product being measured:
a) Instrument configuration (Operating modes)
Density measurement without temperature compensation (TC)
Density measurement with temperature compensation via Pt 100
Density measurement with temperature compensation via current input
Suspension measurement with any carrier liquid
Solids density and liquid density are known
With or without TC
Suspension measurement with water as carrier liquid
- Solids density is known
- with water temperature compensation
the density of the water and
the solids concentration
are used for temperature compensation
Mass flow measurement without TC
Mass flow measurement with TC via Pt 100
b) Special configuration
Back-scatter measurement
c) Selection of measurement mode
Continuous measurement
Discontinuous measurement
- Batch operation via keyboard (press <
run
> button)
- Batch operation via digital input